SPECIFICATIONS
Channel Configuration | 64/128 PR |
Maximum applied voltage (50 Ω load) | PA: 150Vpp (Bipolar) / 75V (Unipolar) UT: 200V |
Maximum PRF | ≤ 30 kHz |
Acquisition | A-Scan/Peak/Conditional data recording |
Acquisition triggered on | Free running, encoder position, external signal |
Digitizing range | 800% |
Max data file size | Limited by hard drive |
Maximum number of reconstruction channels | 128 |
Firing modes | FMC, PWI, Sparse* |
TFM frame size (on board) | 1M points |
TFM frame size (offline) | Unlimited |
Simultaneous FMC channels | 2 |
Max simultaneous TFM frames (on board) | 2 (up to 1M points per frame); 8 (up to 256k points per frame) |
PA connector | 1 IPEX type (2 IPEX with splitter module) |
UT connectors | 4 x LEMO® 00 (8 additional LEMO® 00 with splitter module) |
Data connectivity | Ethernet 5 GBit/second |
Encoder interfaces | 3 quadrature-type |
I/O capability | 12 inputs, 9 outputs |
Automated probe detection | Yes (with Zetec probe ID chip) |
Automated scanner detection | Yes (with Zetec probe ID chip) |
Gain | Up to 124 dB (0.1 dB step), 76 dB Analog / 48 dB Digital |
Input impedance | 50 Ω |
Bandwidth (-3 dB) | PA: 0.5 to 18 MHz, UT: 0.5 to 22.5 MHz |
Data compression | 1, 2, 4, 8, 16 |
Amplitude resolution | 14-bit elementary A-Scan, 16-bit PA |
Max number of samples | 16k |
Max A-scan range | 65k samples |
Measurement gates | 6 + 1 synchronization gate (peak, crossing, autocrossing, homing) |
Parallel PA channels processing | Up to 2 channels * |
Rectification | Digital |
Filtering | Analog / Digital (FIR) |
Digitizing frequency | 100 MHz |
TCG dynamic range | 40 dB |
Specifications listed may change without prior notification.
* – available in a future release