SPECIFICATIONS
QuartZ-Configuration
Phased Array Channels | 32:128 PR |
PA Firing Modes | Up to 32 consecutive elements. Up to two apertures of 16 consecutive elements |
Phased Array Connector | 2 x ZPAC connector (custom ZIF with latch). Allows connecting 2 probes without splitter. Automatic probe recognitions (with Zetec probes). |
UT Channels | 2 channels (in Pulse/Echo or Pitch/Catch configurations |
Pulser
Pulse width | 25 ns to 500 ns |
Pulse amplitude PA (at 50Ω) | 35 V to 100 V |
Pulse amplitude UT (at 50Ω) | 50 V to 200 V |
Data Acquisition
A-scan length | Up to 16,384 points |
Maximum number of focal laws | 1024 |
Real-time averaging | 1, 2, 4, 8, and 16 |
Compression | 1, 2, 4, 8, and 16 |
PRF | Up to 20 kHz |
Parallel Firing | 2 beams |
Maximum number of samples | 16384 |
Measurement Gates | 4 gates + 1 synchronization gate |
Data throughput | Up to 30 MB/s |
Maximum data file | 20 GB |
Digitizing frequency | 25 MHz, 50 MHz or 100 MHz |
Bandwidth (at -3 dB) | 500 kHz to 18 MHz |
Summed data amplitude resolution | 16 bits |
Filters | Analog / Digital band-pass, high-pass and low-pass |
Gain setting range PA | 70 dB (analog) + 30 dB (digital) |
Gain setting range UT | 70 dB (analog) + 24 dB (digital) |
Interfacing
Data Interfaces | Ethernet 1000 Base-T |
Encoder | 2 axes (quadrature, clock direction) |
Housing
Size (H × W × D) | 420x490x90mm (16.5 x 19.3 x 3.5 in) |
Weight | 8.34 Kg |
Air Intake | No |
General specifications
Voltage | 120 VAC or 240 VAC |
Frequency | 50 Hz or 60 Hz |
Maximum power | 75 VA |