search
Menu
 

QuartZ

SPECIFICATIONS

QuartZ-Configuration

Phased Array Channels 32:128 PR
PA Firing Modes Up to 32 consecutive elements. Up to two apertures of 16 consecutive elements
Phased Array Connector 2 x ZPAC connector (custom ZIF with latch). Allows connecting 2 probes without splitter. Automatic probe recognitions (with Zetec probes).
UT Channels 2 channels (in Pulse/Echo or Pitch/Catch configurations

Pulser

Pulse width 25 ns to 500 ns
Pulse amplitude PA (at 50Ω) 35 V to 100 V
Pulse amplitude UT (at 50Ω) 50 V to 200 V

Data Acquisition

A-scan length Up to 16,384 points
Maximum number of focal laws 1024
Real-time averaging 1, 2, 4, 8, and 16
Compression 1, 2, 4, 8, and 16
PRF Up to 20 kHz
Parallel Firing 2 beams
Maximum number of samples 16384
Measurement Gates 4 gates + 1 synchronization gate
Data throughput Up to 30 MB/s
Maximum data file 20 GB
Digitizing frequency 25 MHz, 50 MHz or 100 MHz
Bandwidth (at -3 dB) 500 kHz to 18 MHz
Summed data amplitude resolution 16 bits
Filters Analog / Digital band-pass, high-pass and low-pass
Gain setting range PA 70 dB (analog) + 30 dB (digital)
Gain setting range UT 70 dB (analog) + 24 dB (digital)

Interfacing

Data Interfaces Ethernet 1000 Base-T
Encoder 2 axes (quadrature, clock direction)

Housing

Size (H × W × D) 420x490x90mm (16.5 x 19.3 x 3.5 in)
Weight 8.34 Kg
Air Intake No

General specifications

Voltage 120 VAC or 240 VAC
Frequency 50 Hz or 60 Hz
Maximum power 75 VA